Zoom AP+4 4401A Manual de usuario

Busca en linea o descarga Manual de usuario para Redes Zoom AP+4 4401A. 4401 Oxy-Trace v4 - Barben Analyzer Technology Manual de usuario

  • Descarga
  • Añadir a mis manuales
  • Imprimir

Indice de contenidos

Pagina 1 - 4401 Oxy-Trace v4

1 Instruction Manual Natural Gas Measurement 4401 Oxy-Trace v4 Fiber-optic trace oxygen meter Manual Revision 1

Pagina 2 - Natural Gas

10 10 PROTECTIVEEARTH0-10 VDC(TEMP,PRESSURE) 8 = V1+12 = V1-16 = V2+15 = V2-TEMPERATURE(PT1000)5 = U+6 = U-4 = GNDRS232 INTERFACE AUX DC OUTPU

Pagina 3

11 11 7.0 Required Basic Equipment • Oxygen meter 4401 Oxy-trace NG • Line adapter (110/220 Vac, Optional 12/24 Vdc) • Temperature sensor Pt 1

Pagina 4 - Congratulations!

12 12 8.0 Optical Oxygen Sensors Sensor Characteristic of Oxygen-Sensitive Sensors The operation principle of the optical oxygen sensor is based o

Pagina 5 - 2.0) Safety Guidelines

13 13 Optical isolation (Optional) Optical isolated sensor tips are required in areas where the sensor is not placed in a flow-through fitting, if

Pagina 6 - “Warning”

14 14 Trace Oxygen DP Probe This oxygen sensor consists of a pol ymer optical fiber (POF) with a polished distal tip which is coated with a plan

Pagina 7

15 15 10.0 Configuration of the 4401 Oxy-trace v4 10.1 Powering up the Instrument Prior to powering up the 4401 Oxy refer to Diagram 1 (Page 8) for

Pagina 8

16 16 Time ofDay ClockOxygenConcentrationTemperatureStatusSensorAmplitudeSamplingRateOxygenUnits

Pagina 9 - Wiring Label 1 (VAC Version)

17 17 10.2 Sensor Selection Before starting the measurement you have to choose the correct sensor type. To set or change the sensor type you have

Pagina 10 - Terminal Board Connections

18 18 MeasurementSampling RateTemperaturePressureSignal AverageOxygen UnitsBack EnterSig. Intensity Use the up and down buttons to select. Pres

Pagina 11

19 19 Oxygen units: Press edit to change the oxygen unit displayed. Use the up and down buttons to select the respective unit and confirm by pressin

Pagina 12 - 8.0 Optical Oxygen Sensors

2 Instruction Manual 4401 Oxy-trace v4 Natural Gas PC Software Version 4401TRACEv1.14 October 2006 Specification: Stand-alone one-channel fib

Pagina 13 -

20 20 10.5 Configuration Menu In this menu you can set most of your instrument parameters ConfigurationStatusAnalog OutAnalog InClock/DateDispl

Pagina 14 - Trace Oxygen DP Probe

21 21 Voltage U1/U2: Select the parameter you want to activate as a Voltage analog output for port 1 and port 2 by using the arrow keys. The follow

Pagina 15

22 22 Clock/Date: Use + and next button to adjust clock and date and confirm with SET Display: Changes the b acklight illumination of the display

Pagina 16 - ALTERNATE DISPLAY VIEW

23 23 Press Calibrate to enter into the calibration menu and choose the calibration mode: a) Temperature ON: Calibration with the actual tempe

Pagina 17 - 10.2 Sensor Selection

24 24 8) Turn off the Nitrogen test gas and disconnect from the sensor. Reconnect the “Span” calibration test gas (100 ppm oxygen) and turn on th

Pagina 18

25 25 5. If the right com port is adjusted this information window disappears within a few seconds. If the wrong com port is adjusted you are ask

Pagina 19

26 26 12.2.1 Menu Bar File Charts Display Æ Exit Æ Oxygen Æ Zoom Æ AutoScaleY1 Æ Undo Zoom Æ Phase Æ Amplitude Æ Clear Charts Æ Tem

Pagina 20 - 10.5.1 Analog out:

27 27 12.2.2 Display Zoom: AutoScaleY1 is the default setting. AutoScaleY1 means that the y-axis is scaled automatically. Undo Zoom: The original

Pagina 21 - 10.5.2 Analog In:

28 28 12.2.3 Instrument Information Software Information Barben Analyzer Technology LLC 5200 Convair Drive Carson City, Nevada 89706 Version 8

Pagina 22 - Dianostics and Test

29 29 12.3 LED-Intensity With the current of the LED you can adjust the amount of light illuminating the sensor spot. You can choose between an ‘

Pagina 23

3 Table of Contents Page 1. Preface…………………………………………………………………………………….. 4 2. Safety Guidelines of the 4401 Oxy-NG………………………………

Pagina 24

30 30 Advanced: Click the ‘Advanced’ button to change the LED current manually. Values between 10 and 100 % are possible. After clicking the ‘conf

Pagina 25

31 31 12.4 Oxygen Unit Here you can adjust the desired oxygen unit. Tables and formulas for th e calculation of different concentration scales are

Pagina 26 - 12.2.1 Menu Bar

32 32 Example 1: Using the values given in note 1 and 2, the temperature correction upon the input voltage is: By input voltage = 5V, the i

Pagina 27 - 12.2.2 Display

33 33 Control buttons: The way to start a measurement is (A) Calibration of the mini-sensor with the Calibration Assistant (B) Start Measurement w

Pagina 28 - Software Information

34 34 In this window you can choose the measurement settings: In the ‘Sampling Rate’ window you can select the desired measurement mode with a d

Pagina 29 - 12.3 LED-Intensity

35 35 12.7.2 Temperature Compensated Oxygen Measurements: If you want to measure with temperature compensation, click the ’ on’ button. Please ens

Pagina 30

36 36 To store the data of you r measurement click in t he ‘Logging Setup’ the ‘Measure & Log’ item and click th e button ‘Choose File’. Here y

Pagina 31

37 37 12.7.5 Warning Lights: At the right bottom of t he window you can fin d the amplit ude, phase angle an d three warning light s. The warning

Pagina 32 - 12.7 Control Bar

38 38 sw1: Displays the serial port which is used for communication of the 4401 Oxy-trace v4 device with the PC sw2: Displays the file name in w

Pagina 33

39 39 The following rows, separated by semicolons, list the measuring data. The first two rows contain the date and time, the th ird the log-time in

Pagina 34

4 13.8 Correction for Air Pressure…………………………………………………………..41 14.0 Technical Data…………………………………………………………………………….42 14.1 General Data……………

Pagina 35

40 40 Concentrations units are calculated from % air-saturation by the software. Consequently, changes in the actual atmospheric pressure have no e

Pagina 36

41 41 2025303540455055600 5 10 15 20 25time [h]phase angle [°]100 % air-saturation0 % air-saturation Photo-decomposition test of BOS2, continuously

Pagina 37 - Status Bar

42 42 14.0 Technical Data 14.1 General Data MODES BOS2 oxygen sensor range: 0 - 250 % air-saturation BOS1 oxygen sensor range: 0 - 20 % air-sa

Pagina 38

43 43 Analog Output and Input channels 4401 Oxy-trace v4 – instrument version is supplied with a four independent, programmable 12 bit analog I/O c

Pagina 39

44 44 15.0 Technical Notes POWER 4401 Oxy-trace is available in power from 80- 253 Vac or optional 9 – 24 Vdc. The 4401 Oxy ca n be powered by

Pagina 40

45 45 16.0 Enclosure Drawing

Pagina 41 - 0 % air-saturation

46 46 17.0 Parts List: BOS 3 Trace Sensor DP-BOS3-L5-ST5-NOP-SSJ Power Supply (110 Vac) B5400-1001 Board Set B3600-0004 Calibration Span G

Pagina 42 - 14.0 Technical Data

47 47 19.0 Error Codes: bit 0 = ADC overflow bit 1 = Ambient light bit 2 = no oxygen sensor bit 3 = no temperature sensor bit 4 = current output

Pagina 43

5 5 2.0) Safety Guidelines PLEASE READ THESE INSTRUCTIONS CAREFULLY BEFORE WORKING WITH THIS INSTRUMENT! This device has left our factory after

Pagina 44 - Operation Notes

6 6 The 4401 Oxy-trace-NG may only be operated by qualified personal. This measuring instrument was developed for use in the natural gas industry.

Pagina 45 - 16.0 Enclosure Drawing

7 7 3.0 Description of the 4401 Oxy-trace NG v4 Device The 4401 Oxy-trace NG is a p recision, temperature compensated, stand-alone oxygen meter

Pagina 46 - Trouble Shooting Guide

8 8 4.0 General Instructions: Warm-Up Time The warm up time of the electronic and opto-electronic co mponents of the 4401 Oxy-trace NG is 5 minute

Pagina 47 - 19.0 Error Codes:

9 9 6.0 Terminal Layout and Description Terminal Pin Connection DESCRIPTION FUNCTION SMA Connector SMA fiber connector Connect the fiber optic oxy