External Printer
You can use an external printer if i t is s
upported by the pulse generator
software applicatio n. Use only compatible external printers that have been
tested and qualified for use. R efer to the instructions to connect the cable
("Preparing the PRM for Use " on page 6).
PDF reports generated from saved patient data may be sent to an exte rnal
printer using the Patien t Data Man
agement utility. The utility is accessible from
the Pa tient Data Management button on the PRM startup screen.
External VGA Monitor
You can use an external VGA monito r that can synchronize to any horizontal
scan frequency.
NOTE: Equipment connected to the external connections must compl y with
applicable standards (e.g., IE
C/EN 60950-1 for data processing equipment
and IEC/EN 60601-1 for medical equipment). Anyone connecting such
cables or accessories to the PRM may be configuring a medical system and
is responsible to ensure that t
he system complies with t he requirements of
IEC/EN 60601-1, Clause 16 for medical electrical systems.
Warranty Information
A limited warranty is packaged with the PRM. Unless otherwise agreed, the
PRM remain s the prope rty of B
oston Scientific and Boston Scientificmust
perform all necessary servicing and repair work. For additional copies, contact
Boston Scientific using the information on the back cover.
PREPARING THE PRM FOR USE
Make necessary external connections as directed bel ow.
PreparetheTelemetryWand
Prepare the 6577 Telemetry Wand for the sterile field by following the
sterilization procedures below, or by enclosing t he wand in a sterile surgical
sheath.
CAUTION: The Mod el 6577 Telemetry Wand is shipped nonsterile. If the
telemetrywandistobeusedinasterilefield, it must be active ly sterilized
before use or enclosed in a disposable sterile surgical sheath during use.
Either ethy lene oxide (EO) or steam m ay be used for active sterilization. Foll ow
the cleaning instru ctions ("Cleaning the PRM and Accesso ries" on page 23)
before beginning the sterilization process.
CAUTION: Remove the telemetry wand from all packaging material before
sterilizing it.
Ethylene oxi de (EO) me th od: Fo llow the reco mmendations of the EO
sterilization equipment man ufacturer and al low the specified aeration time
to fully elapse prior to use.
Steam method: Follow customary autoclave procedures for wrapped goods
and limit temperature to 132°C (-0°C, + 5°C), 270°F (-0°F, +9° F).
6