Table 9. PRM Nom inal S pecifications (continued)
Characteristic
Nominal
Bandwidth < 1 MHz
Modulation
ASK/OOK
Effective radiated power
-1.3 dBm
Internal prin ter
Paper type Thermosensitive
Paper width
110 mm (4 in)
Chart speed 10, 25, 50, 100 m m/sec
Electrical Safety Testing-Referen
ce fo r
testing according to the IEC 60601
b
Earth resistance
≤ 100 mΩ
Earth leakage curren t
≤ 5 mA No rmal Condition (NC), ≤ 10 mA
Single Fault Condition (SFC)
Patient leakage current wand
≤ 10 µA Normal Condition (NC) and ≤ 5
0
µA Single Fault Condition (SFC) (m
ains
on applied parts)
Patient leakage current ECG ≤ 100 µ A Normal Condition (NC), ≤ 5
00
µA Single Fault Condition (SFC) (
mains
on applied parts)
Electrical Safety Testing-Refe
rence
for testing according to the IE
C 62353
(Installation, Maintenance
, Repair)
ab
Earthbond testing (Gr oundbo
nd testing)
≤ 300 mΩ including power cable
not
exceeding 3 meters
Equipment leakage — direct m etho d
ECG (BF) and Wand (CF) : ≤ 500 µA
Patient leakage current — direct method
ECG (BF) ≤ 500 µA, Wand (CF): ≤ 5
0 µA
Safety Features
Defibrillator protection
to 5000 V 400 J
a. For questions regarding oper
ation or repair o f the PRM, contact Boston Scientificusingthe
information on the back cover. The PRM must be s erviced by Boston Scientific personnel only.
b. After successfully completing safety testing, verify the PRM continues to meet the essential
performance as defined in the b
eginning of this manual.
Table 10. ZWT Nominal Specifications
Characteristic
Nominal
Safety classification
Ingress protection r ating: I
PX0
Dimensions
17.6 cm (6.9 in) wide, 17.3 cm (6 .8 in)
high, 4.6 cm (3 in) deep
42